DA Wojtasek, HS Teixidor, AF Govoni and IF Gareen
Department of Radiology, New York Hospital/Cornell Medical Center, NY 10021.
We evaluated the image quality of mammograms made by using a new dual- screen, dual-emulsion film combination (Kodak Min-R Fast screen, T-Mat Mll film) that permits reduction of radiation exposure by approximately 50% when compared with a standard single-screen, single-emulsion film system (Kodak Min-R screen, OM-1 film). This new film has been improved when compared with earlier T-Mat M film, including the introduction of an inert dye to reduce light crossover to essentially 0%. Mammogram pairs made with the dual-emulsion film combination and the standard single-emulsion film combination were obtained in 50 patients otherwise undergoing routine mammography. The image pairs were randomized and evaluated by three radiologists who used a three-point scale (better, same, or worse). Each pair was evaluated with regard to parenchymal contrast, sharpness, and latitude, as well as the number and sharpness of calcifications (n = 19) and sharpness of masses (n = 12) when present. All three observers found the dual-emulsion film combination to be better than or the same as the standard with regard to parenchymal sharpness (94-100%), the number and sharpness of calcifications (98-100%), and sharpness of masses (100%). Two observers found the dual-emulsion film combination to be significantly worse (p less than .05) than the standard with respect to parenchymal contrast (72%, 86%), and all three observers rated it significantly worse for film latitude (14 to 42%). Our results suggest that this new dual- emulsion film combination that allows mammography to be performed with less radiation exposure can be used without loss of image quality.
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |