OBJECTIVE. The aim of this study was to evaluate the image quality and diagnostic efficacy of multidetector high-resolution CT (HRCT) by comparing it with single-detector HRCT and by comparing the noise and artifact of multidetector HRCT images with a phantom.
SUBJECTS AND METHODS. Multidetector HRCT with six parameters and single-detector HRCT were performed on cadaveric lungs. The image quality and diagnostic efficacy of multidetector HRCT were evaluated in comparison with those of single-detector HRCT. A phantom was scanned, and image artifact and noise were investigated.
RESULTS. The image quality of multidetector HRCT with axial 1.25 mm x 4i (four images per gantry rotation) mode was equal to that of single-detector HRCT. The image quality of multidetector HRCT with other modes was worse than that on single-detector HRCT. The diagnostic efficacy of multidetector HRCT with high-quality mode (pitch, 3:1) and axial mode was equal to that of single-detector HRCT. The diagnostic efficacy on multidetector HRCT with high-speed mode (pitch, 6:1) was worse than that on single-detector HRCT. In the phantom study, images made in high-speed mode had strong artifacts. Noise in the axial mode was milder than that in high-speed mode but more severe than that in high-quality mode.
CONCLUSION. The image quality of axial HRCT with multidetector CT is equal to that on single-detector HRCT. Axial HRCT with multidetector CT is appropriate for evaluating subtle lung abnormalities, but high-speed mode is unsuitable. Using the high-quality mode degrades image quality but is still worthwhile.
![]()
CiteULike
Complore
Connotea
Del.icio.us
Digg
Reddit
Technorati What's this?
This article has been cited by other articles:
![]() |
E J CHOI, Y-W OH, S Y HAM, K Y LEE, and E-Y KANG Comparison between coronal reformatted images and direct coronal CT images of the swine lung specimen: assessment of image quality with 64-detector row CT Br. J. Radiol., June 1, 2008; 81(966): 463 - 467. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. D. Dodd, C. A. Souza, and N. L. Muller High-resolution MDCT of pulmonary septic embolism: evaluation of the feeding vessel sign. Am. J. Roentgenol., September 1, 2006; 187(3): 623 - 629. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. D. Dodd, C. A. Souza, and N. L. Muller Conventional high-resolution CT versus helical high-resolution MDCT in the detection of bronchiectasis. Am. J. Roentgenol., August 1, 2006; 187(2): 414 - 420. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Beigelman-Aubry, C. Hill, A. Guibal, J. Savatovsky, and P. A. Grenier Multi-Detector Row CT and Postprocessing Techniques in the Assessment of Diffuse Lung Disease RadioGraphics, November 1, 2005; 25(6): 1639 - 1652. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. Studler, T. Gluecker, G. Bongartz, J. Roth, and W. Steinbrich Image Quality from High-Resolution CT of the Lung: Comparison of Axial Scans and of Sections Reconstructed from Volumetric Data Acquired Using MDCT Am. J. Roentgenol., September 1, 2005; 185(3): 602 - 607. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. A. Leswick, S. T. Webster, B. A. Wilcox, and D. A. Fladeland Radiation Cost of Helical High-Resolution Chest CT Am. J. Roentgenol., March 1, 2005; 184(3): 742 - 745. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Coche, F. Verschuren, A. Keyeux, P. Goffette, L. Goncette, P. Hainaut, F. Hammer, E. Lavenne, F. Zech, P. Meert, et al. Diagnosis of Acute Pulmonary Embolism in Outpatients: Comparison of Thin-Collimation Multi-Detector Row Spiral CT and Planar Ventilation-Perfusion Scintigraphy Radiology, December 1, 2003; 229(3): 757 - 765. [Abstract] [Full Text] [PDF] |
||||
![]() |
O. Honda, T. Johkoh, S. Yamamoto, M. Koyama, N. Tomiyama, T. Kozuka, S. Hamada, N. Mihara, H. Nakamura, and N. L. Muller Comparison of Quality of Multiplanar Reconstructions and Direct Coronal Multidetector CT Scans of the Lung Am. J. Roentgenol., October 1, 2002; 179(4): 875 - 879. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |