AJR Get Involved! Join ARRS Today
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Lanners, D. M.
Right arrow Articles by Felmlee, J. P.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Lanners, D. M.
Right arrow Articles by Felmlee, J. P.
Social Bookmarking
 Add to CiteULike   Add to Complore   Add to Connotea   Add to Del.icio.us   Add to Digg   Add to Reddit   Add to Technorati  
What's this?
AJR 2004; 183:825-829
© American Roentgen Ray Society


Breast Imaging

Safety and MRI Artifact Evaluation at 1.5 T of Metallic Mounting Sheath of a Marking Clip Inadvertently Deployed at Stereotactic Biopsy

Diana M. Lanners1, Kimberly K. Amrami1, Renee S. Jonsgaard1, John J. Gisvold1 and Joel P. Felmlee1

1 All authors: Department of Radiology, Mayo Clinic, 201 1st St. SW, Rochester, MN 55905.

Abstract

OBJECTIVE. The purpose of our study was to assess the MRI risk factors for the patient after an unusual case of inadvertent deployment of the mounting sheath of a metallic clip used at stereotactic breast biopsy.

MATERIALS AND METHODS. We evaluated the materials for ferromagnetic properties, heating, and artifacts.

RESULTS. Our analysis showed significant deflection and prominent susceptibility artifacts of the sheath at 1.5 T, although the clip itself showed no deflection and only minimal artifact. Our study shows that a device and its delivery apparatus may have significantly different ferromagnetic properties.

CONCLUSION. In case of inadvertent deployment of the sheath, a delay of 6–8 weeks before MRI is recommended as a conservative approach to ensure tissue ingrowth and to minimize the chance of harm to the patient. In the case of metallic clips used for breast biopsy, caution is warranted when a portion of the device unintended for placement is introduced into the breast.