1 From the Department of Radiology, University of Colorado Medical Center, Denver, Colorado
Radiothermoluminescent measurements have been made of Y90 beta depth dose from a commercially available Sr90-Y90 beta applicator. Good agreement of the resulting beta depth dose curve with those obtained by film and extrapolation chamber techniques of measurement suggests the usefulness of radiothermoluminescent dosimetry as a rapid and convenient approach to the measurement of relative absorbed dose.