Abstract
|
|
|---|
SUBJECTS AND METHODS. One hundred twenty patients underwent MDCT of the trunk (60 thoracic, 60 abdominal) with an attenuation-based online tube current modulation. Consecutive acquisitions at standard and two lower effective mAs presets were obtained in each patient. Mean percentage effective mAs reductions were compared for each effective mAs preset, taking into account sex and body mass index.
RESULTS. Mean effective mAs reduction was 16.9% and 20.0% for the chest and the abdomen, respectively. Mean percentage effective mAs reductions were found to be significantly different for sex (chest, p = 0.003; abdomen, p = 0.002) but not significantly different for the different effective mAs presets or body mass index.
CONCLUSION. Attenuation-based online tube current modulation used with MDCT should be considered as a secondary tool of radiation dose reduction because it saves as much as 20% of the radiation dose on the adult trunk, regardless of initial mAs preset. However, initial decreases of mAs presets by the physician should be considered the primary tool for radiation dose reduction.
|
|
|---|
Various strategies for decreasing the dose, including increased pitch and lowered milliampere-second (mAs) presets (i.e., low-dose CT), have been recommended [4, 5]. Because attenuation of X-rays in the trunk is inhomogeneous and varies with the projection angle, the tube current can be modulated during tube rotation as a function of the projection angle [6]. Applied to standard-dose single-detector CT, attenuation-based online tube current modulation provides approximately a 25% dose reduction with equal or even improved image quality, more homogeneous noise, and fewer attenuation artifacts than with constant tube current [710].
Because this modulation has not yet been investigated using MDCT or low-dose CT, the aim of our study was to quantify the relative reduction in radiation dose achieved by applying this algorithm to MDCT and as a function of effective mAs presets. Because the image quality and the diagnostic accuracy of low-dose CT are also determined by the patient's body habitus [4, 5], we investigated the possible influence of body mass index (BMI) on dose reduction [11]. Finally, because the distribution of body mass may depend on sex, we also investigated the possible influence of BMI on dose reduction.
|
|
|---|
|
CT Examinations
CT scans were obtained using a commercially available helical scanner
(Somatom Plus Volume Zoom, Siemens Medical Systems, Forchheim, Germany) with
four rows of detectors. Patients were examined while in the supine position. A
52-cm scout view was first obtained at 80 kV and 50 mA, followed by three
acquisitions at different effective mAs presets that are summarized in
Table 1. As defined by Mahesh
et al. [12], effective mAs
corresponds to mAs divided by the pitch, whereas the pitch is defined by
Silverman et al. [13] as the
ratio between the table feed per rotation and the X-ray beam width.
|
For the three consecutive acquisitions, the scanned region ranged from the pulmonary apex to the costodiaphragmatic sulci for the chest and from the top of the liver to the symphysis pubis for the abdomen. The effective mAs presets were chosen to reach a total absorbed doseexpressed in weighted CT dose indexnot higher than the reference doses recommended by the European Guidelines on Quality Criteria for Computed Tomography [14] and not exceeding 15.7 and 17.0 mGy for thoracic and abdominal examinations, respectively. The weighted CT dose index used in this study has already been corrected for the pitch in helical and axial scanning. It is equivalent to the new volume CT dose index, as recently introduced by the International Electrotechnical Commission [15].
Chest examinations were not enhanced, whereas abdominal examinations were performed after IV administration of 120 mL of contrast medium with 350 mg of iodine per milliliter (iobitridol, Xenetix 350, Guerbet, Roissy, France) injected at 2 mL/sec after a 70-sec start delay.
Tube Current Modulation
Noise in CT scans varies proportionately to the square root of the applied
dosethat is, proportionately to the square root of the mAs product if
the tube current is kept constant. Noise in CT scans is dominated by those
projections in which attenuation is the highest. For a homogeneous object with
a circular cross section, attenuation is constant over all projections, and
all measured values contribute equally. However, for a nonhomogeneous object
with a noncircular cross section, attenuation varies stronglysometimes
by more than three orders of magnitude
[6,
10]. Noise in the data
measured from high-attenuation projections (i.e., lateral direction) greatly
influences noise level in the CT scans. This means that the dose for
projections with relatively low attenuation (i.e., anteroposterior direction)
can be reduced substantially without a measurable increase in image noise
[6,
10]. The tube current should
thus be decreased as a function of rotation angle whenever attenuation is
low.
The commercially available current modulation software (Care Dose, Siemens Medical Systems) used during all acquisitions in this study is characterized by online monitoring of the attenuation and subsequent tuning of the tube current as a function of the projection angle with a delay of 360°. For projections with low attenuation, the maximal reduction of the tube current is 90%. For each acquisition, the CT unit calculates the arithmetic average effective mAs throughout the duration of the exposure. The mean effective mAs, as displayed on the CT scans, was recorded for further calculations.
Statistical Analysis
Mean percentage effective mAs reductions were compared for the three
effective mAs presets taking into account sex and BMI for the thoracic and
abdominal examinations, respectively. Analyses of variance for repeated
measures, corresponding to each of the three different effective mAs presets,
were performed with sex as the intersubjects factor and BMI as the covariate.
The two-way interactions between dose, sex, and BMI were also investigated.
Statistical significance for all tests was set at a p value of less
than 0.05.
|
|
|---|
|
|
When we compared the sex of patients, mean percentage effective mAs reductions were significantly higher in men than in women for the three effective mAs presets applied to the chest (p = 0.018, 0.001, and 0.004 for 20, 40, and 80 effective mAs presets, respectively) and were significantly lower in men than in women for the three mAs presets applied to the abdomen (p = 0.001, 0.001, and 0.004 for 30, 50, and 100 effective mAs presets, respectively).
This difference according to sex was found to be dependent on BMI for the chest (p = 0.001, 0.015, and 0.004 for 20, 40, and 80 effective mAs presets, respectively) but not for the abdomen (p = 0.388, 0.171, and 0.259 for 30, 50, and 100 effective mAs presets, respectively). The difference according to sex in thoracic examinations reached high statistical significance in the normal BMI subgroup of patients only (p = 0.001, < 0.001, and < 0.001 for 20, 40, and 80 effective mAs presets, respectively).
|
|
|---|
In our study, we considered 80 and 100 effective mAs as standard presets [1719], corresponding to mAs values of 160 and 150 mAs and to tube currents of 320 and 300 mA for the chest and the abdomen, respectively. These presets were set at levels in which no degradation of image quality had been observed in comparison with doubled values of mAs pressets as shown by Ravenel et al. [17] in the chest and by Kalra et al. [18] in the abdomen. The lowest effective mAs presets tested in our study are the ones currently used in screening programs [20, 21]. We have also considered the middle effective mAs preset used in our study as a possible low-dose preset suitable for patients who are overweight.
The effective mAs reduction obtained with attenuation-based online tube current modulation is independent of the effective mAs preset. Because this modulation saves up to only 20% of the dose, it cannot by itself save up to 90% of the dose as the so-called "low-dose" CT protocols do by lowering the effective mAs preset [4, 5, 1921]. Nevertheless, when applied to low-dose presets, the attenuation-based online tube current modulation provides a 1520% supplementary dose reduction [5, 20, 21].
The effective mAs reduction obtained with attenuation-based online tube current modulation is independent of BMI, indicating that normal weight and obese patients have similar ratios in attenuation between anteroposterior and lateral projections. In fact, tube current modulation does not adapt effective mAs settings to the body's diameter nor to the total absorption of X-rays by the patient's body. If CT parameters are maintained constant, independently of body size, the energy delivered to small individuals is lower than that delivered to larger individuals, but the effective dose delivered to sensitive organs is higher in smaller individuals [22, 23]. As suggested by Kalra et al. [18], the effective mAs preset should thus be adapted to the patient's body size (i.e., BMI) before applying attenuation-based online tube current modulation. This recommendation applies to pediatric patients as well.
The effective mAs reduction obtained with attenuation-based online tube current modulation varies also with sex. The mAs reduction was found to be 2% higher in men than in women for the chest and 3% higher in women than in men for the abdomen. However, even if statistically significant, these differences are clinically small. They indicate that men and women have subtle differences in attenuation between anteroposterior and lateral projections. In the chest CT only, this difference is related to BMI, probably as a result of a woman's breasts. Interestingly, sex differences (because of breasts) were seen only in patients of normal weight.
The software release used in our study to modulate online tube current was designed only as a function of the projection angle with a delay of 360°. It was not designed to tune the mAs as a function of the table position (z-axis). Therefore, the tube current was not modulated according to differences in absorption along the scanned region. Because Itoh et al. [24] have reported that the detection of lung nodules requires a tube current that may vary by 50% from the shoulders to the mid lung zones, the attenuation-based online tube current modulation along the cephalocaudal axis could be a technical advance able to reduce the radiation dose more adequately [25].
In conclusion, attenuation-based online tube current modulation used with MDCT reduces radiation dose by 1520% in all patients, regardless of the initial effective mAs preset. As a consequence, attenuation-based online tube current modulation does not replace the reduction in mAs presets but should be considered as a supplementary tool to decrease the radiation dose.
|
|
|---|
This article has been cited by other articles:
![]() |
J. R. Mayo Radiation Dose Issues in Longitudinal Studies Involving Computed Tomography Proceedings of the ATS, December 15, 2008; 5(9): 934 - 939. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Roggenland, S. A. Peters, S. P. Lemburg, T. Holland-Letz, V. Nicolas, and C. M. Heyer CT Angiography in Suspected Pulmonary Embolism: Impact of Patient Characteristics and Different Venous Lines on Vessel Enhancement and Image Quality Am. J. Roentgenol., June 1, 2008; 190(6): W351 - W359. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Matsubara, K. Koshida, M. Suzuki, H. Tsujii, T. Yamamoto, and O. Matsui Comparison between 3-D and z-axis automatic tube current modulation technique in multidetector-row CT Radiat Prot Dosimetry, January 1, 2008; 128(1): 106 - 111. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Coursey, D. P. Frush, T. Yoshizumi, G. Toncheva, G. Nguyen, and S. B. Greenberg Pediatric Chest MDCT Using Tube Current Modulation: Effect on Radiation Dose with Breast Shielding Am. J. Roentgenol., January 1, 2008; 190(1): W54 - W61. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. D. MacKenzie, J. Nazario-Larrieu, T. Cai, M. S. Ledbetter, M. A. Duran-Mendicuti, P. F. Judy, and F. J. Rybicki Reduced-Dose CT: Effect on Reader Evaluation in Detection of Pulmonary Embolism Am. J. Roentgenol., December 1, 2007; 189(6): 1371 - 1379. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. A. Jaffe, A. M. Gaca, S. Delaney, T. T. Yoshizumi, G. Toncheva, G. Nguyen, and D. P. Frush Radiation Doses from Small-Bowel Follow-Through and Abdominopelvic MDCT in Crohn's Disease Am. J. Roentgenol., November 1, 2007; 189(5): 1015 - 1022. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. M. Kanal, B. K. Stewart, O. Kolokythas, and W. P. Shuman Impact of Operator-Selected Image Noise Index and Reconstruction Slice Thickness on Patient Radiation Dose in 64-MDCT Am. J. Roentgenol., July 1, 2007; 189(1): 219 - 225. [Abstract] [Full Text] [PDF] |
||||
![]() |
A J Van Der Molen, W J H Veldkamp, and J Geleijns 16-slice CT: achievable effective doses of common protocols in comparison with recent CT dose surveys Br. J. Radiol., April 1, 2007; 80(952): 248 - 255. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Graser, B. J. Wintersperger, C. Suess, M. F. Reiser, and C. R. Becker Dose reduction and image quality in MDCT colonography using tube current modulation. Am. J. Roentgenol., September 1, 2006; 187(3): 695 - 701. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Coche, S. Vynckier, and M. Octave-Prignot Pulmonary Embolism: Radiation Dose with Multi-Detector Row CT and Digital Angiography for Diagnosis Radiology, September 1, 2006; 240(3): 690 - 697. [Abstract] [Full Text] [PDF] |
||||
![]() |
G R Iball, D S Brettle, and A C Moore Assessment of tube current modulation in pelvic CT Br. J. Radiol., January 1, 2006; 79(937): 62 - 70. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Campbell, M. K. Kalra, S. Rizzo, M. M. Maher, and J.-A. Shepard Scanning Beyond Anatomic Limits of the Thorax in Chest CT: Findings, Radiation Dose, and Automatic Tube Current Modulation Am. J. Roentgenol., December 1, 2005; 185(6): 1525 - 1530. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. K. Kalra, S. Rizzo, M. M. Maher, E. F. Halpern, T. L. Toth, J.-A. O. Shepard, and S. L. Aquino Chest CT Performed with Z-Axis Modulation: Scanning Protocol and Radiation Dose Radiology, October 1, 2005; 237(1): 303 - 308. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. H. Mulkens, P. Bellinck, M. Baeyaert, D. Ghysen, X. Van Dijck, E. Mussen, C. Venstermans, and J.-L. Termote Use of an Automatic Exposure Control Mechanism for Dose Optimization in Multi-Detector Row CT Examinations: Clinical Evaluation Radiology, October 1, 2005; 237(1): 213 - 223. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Tack, P. Bohy, I. Perlot, V. De Maertelaer, O. Alkeilani, S. Sourtzis, and P. A. Gevenois Suspected Acute Colon Diverticulitis: Imaging with Low-Dose Unenhanced Multi-Detector Row CT Radiology, October 1, 2005; 237(1): 189 - 196. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. C. O'Daniel, D. M. Stevens, and D. D. Cody Reducing Radiation Exposure from Survey CT Scans Am. J. Roentgenol., August 1, 2005; 185(2): 509 - 515. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Das, A. H. Mahnken, G. Muhlenbruch, A. Stargardt, C. Weiss, D.-A. Sennst, T. G. Flohr, R. W. Gunther, and J. E. Wildberger Individually Adapted Examination Protocols for Reduction of Radiation Exposure for 16-MDCT Chest Examinations Am. J. Roentgenol., May 1, 2005; 184(5): 1437 - 1443. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. K. Kalra, M. M. Maher, R. V. D'Souza, S. Rizzo, E. F. Halpern, M. A. Blake, and S. Saini Detection of Urinary Tract Stones at Low-Radiation-Dose CT with Z-Axis Automatic Tube Current Modulation: Phantom and Clinical Studies Radiology, May 1, 2005; 235(2): 523 - 529. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. A. Leswick, S. T. Webster, B. A. Wilcox, and D. A. Fladeland Radiation Cost of Helical High-Resolution Chest CT Am. J. Roentgenol., March 1, 2005; 184(3): 742 - 745. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. M. R. Rizzo, M. K. Kalra, M. M. Maher, M. A. Blake, T. L. Toth, and S. Saini Do Metallic Endoprostheses Increase Radiation Dose Associated with Automatic Tube-Current Modulation in Abdominal-Pelvic MDCT? A Phantom and Patient Study Am. J. Roentgenol., February 1, 2005; 184(2): 491 - 496. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. K. Kalra, M. M. Maher, R. S. Kamath, T. Horiuchi, T. L. Toth, E. F. Halpern, and S. Saini Sixteen-Detector Row CT of Abdomen and Pelvis: Study for Optimization of Z-Axis Modulation Technique Performed in 153 Patients Radiology, October 1, 2004; 233(1): 241 - 249. [Abstract] [Full Text] [PDF] |
||||
![]() |
J.-F. Paul, H. T. Abada, and A. B. Sigal-Cinqualbre Automatic Dose Reduction Should Not Mask Needs for Individual Dose Reduction Radiology, October 1, 2004; 233(1): 297 - 297. [Full Text] [PDF] |
||||
![]() |
M. K. Kalra, M. M. Maher, T. L. Toth, R. S. Kamath, E. F. Halpern, and S. Saini Comparison of Z-Axis Automatic Tube Current Modulation Technique with Fixed Tube Current CT Scanning of Abdomen and Pelvis Radiology, August 1, 2004; 232(2): 347 - 353. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Beyer, G. Antoch, S. Muller, T. Egelhof, L. S. Freudenberg, J. Debatin, and A. Bockisch Acquisition Protocol Considerations for Combined PET/CT Imaging J. Nucl. Med., January 1, 2004; 45(90010): 25S - 35. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |