|
|
||||||||
Original Research |
1 Institute of Diagnostic Radiology, University Hospital Zurich, Raemistrasse
100, 8091 Zurich, Switzerland.
2 Siemens Medical Solutions, Forchheim, Germany.
3 Cardiovascular Center, University Hospital, Zurich, Switzerland.
4 Zurich Center for Integrative Human Physiology, Zurich, Switzerland.
Received December 18, 2006;
accepted after revision February 28, 2007.
Address correspondence to H. Alkadhi.
Abstract
|
|
|---|
SUBJECTS AND METHODS. Eighty patients underwent dual-source CT coronary angiography. Thirteen data sets were reconstructed in 5% steps from 20-80% of the R-R interval. Heart rate variability was calculated as SD of mean heart rate. Two independent blinded reviewers assessed the image quality of each segment.
RESULTS. Mean heart rate was 65.3 ± 13.9 (SD) beats per minute (bpm) (range, 35-99 bpm) with a variability of 3.4 ± 4.1 bpm (range, 0.4-17.5 bpm). Image quality was sufficient for diagnosis for 97.8% (1,043/1,066) of arterial segments. No significant correlation was found between mean heart rate and image quality in any segment or any coronary artery. No significant correlation was found between heart rate variability and image quality in any segment, the right coronary artery, or the left anterior descending artery, but there was a significant (p < 0.05) correlation in the left circumflex artery. A significant correlation (p < 0.05) between overall image quality was found for mean and variability of heart rate as shared predictors, the latter having a greater contribution.
CONCLUSION. The overall image quality of dual-source CT coronary angiography is sufficient for diagnosis within a wide range of mean heart rates and variability of heart rates. Only heart rates that are both high and variable significantly deteriorate image quality, but the quality remains adequate for diagnosis.
Keywords: coronary angiography dual-source CT heart rate image quality
|
|
|---|
Dual-source CT is characterized by two X-ray tubes and two corresponding detectors mounted on the gantry with an angular offset of 90° [14]. With a gantry rotation time of 330 milliseconds, this scanner has heart rate-independent temporal resolution of 83 milliseconds with use of a single-segment reconstruction mode. Results of preliminary studies [15, 16] have shown that the quality of images obtained with dual-source CT coronary angiography at a relatively high mean heart rate is sufficient for diagnosis. Moreover, an early experience [17] showed that compared with invasive coronary angiography, dual-source CT coronary angiography had high diagnostic accuracy in the diagnosis of substantial coronary artery stenosis in a patient population whose heart rates were not controlled. The purpose of our study was to systematically evaluate the effect of the relation between mean heart rate and variability of heart rate during scanning on the image quality of dual-source CT coronary angiography.
|
|
|---|
|
Dual-Source CT Protocol
All CT examinations were performed with a dual-source CT scanner (Somatom
Definition, Siemens Medical Solutions). Scanning was conducted in a
craniocaudal direction covering the region from 1 cm caudal to the level of
the tracheal bifurcation to the diaphragm. Scanning parameters were as
follows: detector collimation, 2 x 32 x 0.6 mm; slice collimation,
2 x 64 x 0.6 mm by means of a z-flying focal spot; gantry rotation
time, 330 milliseconds; heart rate-dependent pitch, 0.2-0.43; tube
current-time product, 350 mAs per rotation; tube potential, 120 kV. The
ECG-pulsing technique for radiation dose reduction was used in all patients
with full tube current from 20% to 80% of the R-R interval. With this
protocol, mean dose-length product was 703.4 ± 196.3, and mean CT dose
index was 56.2 ± 16.9.
Two minutes before CT, all patients received a single sublingual 2.5-mg dose of isosorbide dinitrate (Isoket, Schwarz Pharma). Eighty milliliters of nonionic isoosmolar contrast material (iodixanol 320 mg I/mL, Visipaque 320, GE Healthcare) was injected at 5 mL/s and followed by a 30-mL saline flush. Bolus-tracking was performed in the ascending aorta, and image acquisition was started after attainment of a threshold of 100 H with a delay of 5 seconds.
CT Data Postprocessing and Analysis
All images were reconstructed with retrospective ECG gating. A monosegment
reconstruction algorithm consisting of the data from a quarter rotation of
both detectors was used for image reconstruction
[14]. For each patient, 13
data sets were reconstructed in 5% steps from 20% to 80% of the R-R interval
with a slice thickness of 0.75 mm, reconstruction increment of 0.5 mm, and
medium soft-tissue convolution kernel (B26f). The field of view was manually
adjusted to encompass the heart (mean, 156 ± 21 mm; range, 130-182 mm;
image matrix, 512 x 512 pixels). All images were transferred to an
external workstation (Leonardo, Siemens Medical Solutions) equipped with
cardiac postprocessing software (Syngo Circulation, Siemens).
The coronary arteries were classified into 15 segments according to the scheme proposed by the American Heart Association [18]. The right coronary artery (RCA) included segments 1-4, the left main coronary artery and left anterior descending coronary artery (LAD) included segments 5-10, and the left circumflex coronary artery (LCX) included segments 11-15. If present, the intermediate artery was designated segment 16. Coronary artery analysis was performed on all vessels with at least 1-mm luminal diameter at the origin. Images were analyzed and graded by two independent reviewers blinded to mean heart rate and heart rate variability during scanning and using axial source images, multiplanar reformations, and thin-slab maximum intensity projections. The reviewers analyzed each coronary artery segment first at 60% and 70% of the R-R interval. If the images were not of sufficient quality for diagnosis at 60% and 70%, additional data sets were reconstructed in 5% steps within the time interval of full tube current from 20% to 80% to obtain images of diagnostic quality. After finding the reconstruction interval with the best image quality, the reviewers semiquantitatively assessed image quality and degree of motion artifacts on the 4-point Likert scale as follows: 1, excellent, no motion artifacts, clear delineation of the segment; 2, good, minor artifacts, mild blurring of the segment; 3, adequate, moderate artifacts, moderate blurring without structure discontinuity; and 4, not evaluative, doubling or discontinuity in the course of the segment preventing evaluation or vessel structures not differentiable (Figs. 1, 2, 3 and 4). Scores 1-3 were considered diagnostic. In case of disagreement between reviewers, consensus interpretation was appended.
|
|
|
|
Statistical Analysis
Quantitative variables were expressed as mean ± SD and categoric
variables as frequencies or percentages. Interobserver agreement for image
quality assessment was interpreted by use of kappa values. Pearson's
correlation analysis was performed to compare image quality scores of all
segments together and separately for the RCA, LAD, and LCX in each patient
with mean heart rate and with the SD of the mean heart rate (i.e., heart rate
variability) during scanning. To test for colinearity, Pearson's correlation
analysis was performed between mean heart rate and heart rate variability.
Multivariate regression analysis was performed to test for mutual effects of
mean heart rate and heart rate variability on mean image quality score per
patient. Individual contribution to image quality modification was assessed
with standardized beta coefficients. Wilcoxon's two-sample test was performed
to evaluate the effect of ß-blocker medication on heart rate variability
and mean heart rate. Two-tailed p < 0.05 was considered
statistically significant. All statistical analysis was performed with
commercially available software (SPSS 12.0, SPSS).
|
|
|---|
= 0.72).
Table 2 summarizes image
quality scores for all segments and coronary arteries.
|
Effect of Mean Heart Rate on Image Quality
The mean heart rate during scanning was 65.3 ± 13.9 bpm (range,
35-99 bpm). No significant correlation was found between mean heart rate and
mean overall image quality of dual-source CT coronary angiography in any
segment in any patient (r = 0.20, p =not significant [NS])
(Fig. 5). Similarly, no
significant correlation was present between mean heart rate and quality of
images of the RCA (r = 0.11, p = NS), LAD (r =
0.18, p =NS), or LCX (r = 0.19, p =NS).
|
|
Effect of ß-Blockers on Mean Heart Rate, Heart Rate Variability, and Image Quality
There was a significant difference between the mean heart rate of patients
not taking ß-blocker medication (69.0 ± 14.4 bpm) and that of
patients taking ß-blockers (62.2 ± 12.9 bpm). The mean heart rate
was significantly lower among patients taking baseline ß-blocker
medication (p < 0.01). The variability of heart rate among
patients taking ß-blocker medication (3.6 ± 4.2 bpm) was not
significantly different (p = NS) from that of those not taking
ß-blockers (3.3 ± 4.2 bpm). There was no significant difference in
image quality between patients taking ß-blocker medication (1.54 ±
0.38) and those not taking ß-blockers (1.61 ± 0.40, p
=NS).
|
|
|---|
Mean Heart Rate and Image Quality
CT coronary angiographic studies of each coronary artery with 4-MDCT at a
gantry rotation time of 500 milliseconds had significantly decreased image
quality with increasing mean heart rates
[10]. Using 16-MDCT at a
gantry rotation time of 420 milliseconds, Hoffmann et al.
[11] found a significant
negative correlation between overall image quality and mean heart rate.
Studies with 64-MDCT at gantry rotation times of 370 and 330 milliseconds have
shown that CT coronary angiography can be performed with diagnostic image
quality at a wide range of heart rates. Whereas one study
[12] showed a significant
negative relation between overall image quality and mean heart rate, another
study [13] showed the relation
to overall image quality as nonsignificant. The latter study showed a weak
though significant negative correlation between mean heart rate and image
quality only for the LCX. This effect was believed to be caused by the
increased susceptibility of the LCX to shortening of diastole at higher heart
rates and by its higher motion velocity
[20]. Using dual-source CT
with a gantry rotation time of 330 milliseconds, we found no significant
correlation between mean heart rate and the overall image quality for any
coronary segment or for any individual coronary artery. We found no motion
artifacts in images of 54.4% of segments and minor to moderate motion
artifacts in images of 43.4% of coronary segments. Severe artifacts leading to
nondiagnostic image quality were present in images of only 2.2% of the
segments, almost exclusively involving distal segments or side branches.
Although direct comparison of our results with those from a previous 64-MDCT
study [13] is not feasible,
our data indicate that the increased temporal resolution of 83 milliseconds
with dual-source CT allows imaging of the coronary arteries with diagnostic
quality at heart rates up to 100 bpm.
Heart Rate Variability and Image Quality
With interheartbeat variability of the heart rate during scanning, the
ECG-gated image reconstruction technique at a fixed interval does not generate
images in the same cardiac phases. This problem is caused by nonproportional
shortening and prolongation of the cardiac phases at different heart rates
[23]. For images reconstructed
with monosegment techniques, data obtained from slightly different cardiac
phases result in a time shift from image to image, giving reformats a
stair-step appearance. Multisegment algorithms for data reconstruction merge
data from two or more consecutive heartbeats to increase the temporal
resolution [24]. Use of these
algorithms in CT examinations with variable heart rates leads to merging of
data that do not exactly match. Although this technique may reduce the
appearance of stair-step artifacts, blurring of images increases
[12,
13,
24].
The issue of heart rate variability with regard to CT coronary angiography was addressed in a 64-MDCT study by Leschka et al. [13]. By using a two-segment reconstruction algorithm at heart rates greater than 65 bpm, the authors found significant negative correlation between heart rate variability and image quality for the whole coronary artery tree and each coronary artery. This finding is in contrast to our findings with dual-source CT. We found no significant effect of interheartbeat variation on overall image quality. Because the increased temporal resolution of dual-source CT is not likely to be beneficial for improving image quality in scans with variable heart rates, the most probable cause of our findings is the use of a monosegment reconstruction algorithm that does not merge data from adjacent heartbeats in different cardiac phases. Nonetheless, we found a significant coeffect of both heart rate variability and mean heart rate on the image quality of dual-source CT coronary angiography. The contribution of heart rate variability was greater than that of mean heart rate. This finding indicates that the combination of high and irregular heart rates negatively affects the image quality of noninvasive coronary angiography, even with dual-source CT. On the other hand, most of the not-evaluative segments were distal segments or side branches in 11 patients who had a mean heart rate of 66 ± 11 bpm and a heart rate variability of 3 ± 3 bpm, suggesting vessel size but not heart rate is the main limiting factor for image quality. A more proximal segment (mid RCA) was considered not evaluative in only one patient, who had a mean heart rate of 68 bpm but a high heart rate variability of 8.4 bpm during scanning.
ß-Receptor Blockers and Image Quality
Reduction of heart rate with ß-blocker medication has been reported to
improve the image quality of CT coronary angiography
[13,
22,
25] and has been proposed in
most 64-MDCT coronary angiography studies
[2-4,
6,
26]. The main mechanism of
ß-blockers in improving image quality is its ability to reduce heart rate
variability [13]. We found no
significant difference between the heart rate variability of patients taking
ß-blocker medication and that of patients not taking the medication. This
finding is in contrast to the findings of Leschka et al.
[13] and may be explained by
different indications for ß-blocker treatment, individual response to
ß-blocker medication
[27], and by our cohort's
being too small to reveal statistically significant differences between
subgroups. Nevertheless, because overall image quality for all patients was
not related to either mean heart rate or variability of heart rate, our
results suggest that heart rate control with ß-blockers is not necessary
before dual-source CT coronary angiography. This suggestion is supported by
results of a study [17] of the
performance of dual-source CT coronary angiography in the diagnosis of
coronary artery stenosis. That study showed high diagnostic accuracy in
patients not receiving heart rate control. On the other hand, our results
indicate that administration of ß-blockers may be beneficial in improving
the image quality of dual-source CT of patients whose heart rate is both high
and variable.
Study Limitations
We acknowledge the following study limitations. First, image quality
scoring might have been influenced by a subjectivity bias. Second, we did not
analyze the influence of heart rate values on the diagnostic accuracy of
coronary angiography. Twenty-four patients admitted for suspected coronary
artery disease declined invasive coronary angiography because of negative
findings on CT. Therefore, we did not include coronary artery stenosis
detection in our study, which might have resulted in bias due to an incomplete
data set. Finally, we did not investigate the effect of heart rate values on
the image quality of CT coronary angiography with the absolute ECG-gating
technique. This technique has been suggested to be advantageous in patients
with variable heart rates
[28], but there is no
difference between relative and absolute ECG gating in patients with regular
sinus rhythm.
Conclusion
In evaluation of the coronary arteries, dual-source CT has diagnostic image
quality within a wide range of mean and variable heart rates. Neither mean
heart rate nor variability of heart rate has a negative effect on the overall
image quality of dual-source CT coronary angiography. Only heart rates that
are both high and variable deteriorate image quality, but the quality remains
sufficient for diagnosis.
|
|
|---|
This article has been cited by other articles:
![]() |
H. Matsumoto, T. Kondo, S. Watanabe, R. Kikumoto, T. Shimada, Y. Hiraoka, and K. Ueda ECG-Edited Middiastolic Phase Reconstruction Improves Image Quality at 64-MDCT Coronary Angiography of Patients with Atrial Fibrillation Am. J. Roentgenol., December 1, 2008; 191(6): 1659 - 1666. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Stolzmann, H. Scheffel, S. Leschka, A. Plass, S. Baumuller, B. Marincek, and H. Alkadhi Influence of Calcifications on Diagnostic Accuracy of Coronary CT Angiography Using Prospective ECG Triggering Am. J. Roentgenol., December 1, 2008; 191(6): 1684 - 1689. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Stolzmann, S. Leschka, H. Scheffel, T. Krauss, L. Desbiolles, A. Plass, M. Genoni, T. G. Flohr, S. Wildermuth, B. Marincek, et al. Dual-Source CT in Step-and-Shoot Mode: Noninvasive Coronary Angiography with Low Radiation Dose Radiology, October 1, 2008; 249(1): 71 - 80. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Bastarrika, C. N. De Cecco, M. Arraiza, S. Mastrobuoni, J. C. Pueyo, M. Ubilla, and G. Rabago Dual-Source CT for Visualization of the Coronary Arteries in Heart Transplant Patients with High Heart Rates Am. J. Roentgenol., August 1, 2008; 191(2): 448 - 454. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. P. Shuman, K. R. Branch, J. M. May, L. M. Mitsumori, D. W. Lockhart, T. J. Dubinsky, B. H. Warren, and J. H. Caldwell Prospective versus Retrospective ECG Gating for 64-Detector CT of the Coronary Arteries: Comparison of Image Quality and Patient Radiation Dose Radiology, August 1, 2008; 248(2): 431 - 437. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Husmann, I. Valenta, O. Gaemperli, O. Adda, V. Treyer, C. A. Wyss, P. Veit-Haibach, F. Tatsugami, G. K. von Schulthess, and P. A. Kaufmann Feasibility of low-dose coronary CT angiography: first experience with prospective ECG-gating Eur. Heart J., January 2, 2008; 29(2): 191 - 197. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. M. Colletti Coronary CT Angiography Without {beta}-Blockers Am. J. Roentgenol., December 1, 2007; 189(6): 1324 - 1325. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |