DOI:10.2214/AJR.07.2296
AJR 2007; 189:981-988
© American Roentgen Ray Society
Soft and Intermediate Plaques in Coronary Arteries: How Accurately Can We Measure CT Attenuation Using 64-MDCT?
Jun Horiguchi1,
Chikako Fujioka1,
Masao Kiguchi1,
Yun Shen2,
Christian E. Althoff3,4,
Hideya Yamamoto5 and
Katsuhide Ito3
1 Department of Clinical Radiology, Hiroshima University Hospital, 1-2-3,
Kasumi-cho, Minami-ku, Hiroshima, 734-8551, Japan.
2 CT Lab of Great China, GE Healthcare, Mongkok Kowloon, Hong Kong.
3 Department of Radiology, Division of Medical Intelligence and Informatics,
Programs for Applied Biomedicine, Graduate School of Biomedical Sciences,
Hiroshima University, Hiroshima, Japan.
4 Present address: Institute of Radiology, Universitaetsmedizin-Charité
-Berlin, Berlin, Germany.
5 Department of Molecular and Internal Medicine, Division of Clinical Medical
Science, Programs for Applied Biomedicine, Graduate School of Biomedical
Sciences, Hiroshima University, Hiroshima, Japan.
Received November 15, 2006;
accepted after revision May 18, 2007.
Y. Shen is an employee of GE Healthcare.
Address correspondence to J. Horiguchi.
Abstract
OBJECTIVE. The objective of this study was to validate the accuracy
of 64-MDCT densitometry of soft and intermediate plaques.
MATERIALS AND METHODS. Acrylonitrile–butadiene–styrene
resin (47 H) and acrylic (110 H) were used to simulate soft and intermediate
plaques, respectively, in coronary artery models (diameters of 3 and 4 mm).
The variable parameters were heart rate (50, 65, 80, and 95 beats per minute),
reconstruction algorithm (half and segmentation), coronary artery enhancement
(150, 250, 350, and 450 H), CT densitometry site (arterial lumen or center),
shape of plaque (D-shaped, centric, and eccentric), and level of stenosis due
to plaque (25%, 50%, and 75% of arterial diameter). Measured CT attenuation
values of soft and intermediate plaques were compared for different
combinations of parameters. Repeated measures analysis of variance, Wilcoxon's
signed rank, Mann-Whitney U, and Kruskal-Wallis tests were used for
statistical analyses.
RESULTS. For measuring soft plaque, CT densitometry was accurate at
low heart rates with the use of a half reconstruction algorithm (p
< 0.01) on intracoronary artery enhancement of 250 H (p <
0.01). For both soft and intermediate plaques, the densitometry measurements
near the arterial lumen were overestimated and higher than those at the center
(p < 0.01). For plaques that were 50% or more of the arterial
diameter, accurate CT densitometry was possible.
CONCLUSION. Coronary artery enhancement has a significant impact on
64-MDCT densitometry measurements of coronary artery plaques, especially of
soft plaques. A large plaque size, densitometry performed not near the
arterial lumen but at the center of the plaque, intracoronary enhancement of
250 H, and a low heart rate increase the accuracy of plaque densitometry.
Keywords: cardiac CT coronary artery plaque
Introduction
Most patients with acute coronary syndromes present with unstable angina,
acute myocardial infarction, and sudden coronary death. The most common cause
of coronary thrombosis is plaque rupture followed by plaque erosion, whereas
calcified nodules are infrequently associated with erosion
[1]. The term "vulnerable
plaque" was introduced to define plaques susceptible to such ischemic
complications [2]. The
detection and characterization of vulnerable plaques remain difficult tasks
even using invasive techniques such as intravascular sonography
[3], optical coherence
tomography [4], plaque
thermography [5], and
angioscopy [6].
Kopp et al. [7] opened the
way to plaque characterization using MDCT by calculating CT attenuation.
Schroeder et al. [8], in their
analysis of the composition of 34 plaques on 4-MDCT and intracoronary
sonography, found that soft plaques had a mean attenuation (± SD) of 14
± 26 H (range, –42 to 47 H) and intermediate plaques, of 91
± 21 H (range, 61–112 H). Leber et al.
[9], in their analysis of 58
plaques using 16-MDCT, showed that soft plaques had a mean attenuation of 49
± 22 H (range, 14–82 H) and intermediate plaques, of 91 ±
22 H (range, 34–125 H), although the values for soft and intermediate
plaques overlapped. In the most recent study to date of 16-MDCT involving 252
plaques, Pohle et al. [10]
found that the attenuation values of soft (58 ± 43 H) and intermediate
(121 ± 34 H) plaques were statistically different; however, there was
substantial overlap. They therefore concluded that the differentiation of
"vulnerable" from "stable" plaques based on their CT
attenuation is doubtful.
On the other hand, the results of some ex vivo studies indicate that plaque
differentiation may be possible
[11–13].
After their analysis of the composition of 34 plaques on 4-MDCT and
intracoronary sonography [8],
Schroeder et al. [13]
suggested the following MDCT attenuation criteria for differentiation of
plaques: £ 60 H, predominantly lipid-rich plaques; 61–119 H,
intermediate plaques; and
120 H, predominantly calcific plaques.

View larger version (12K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 2 —Graph shows eight heart rate sequences that were programmed
in phantom: sequence 1, 50 beats per minute (bpm); sequence 2, 65 bpm;
sequence 3, 80 bpm; sequence 4, 95 bpm; sequence 5, 50 bpm with shifting;
sequence 6, 65 bpm with shifting; sequence 7, 80 bpm with shifting; and
sequence 8, 95 bpm with shifting.
|
|
Using static cardiac phantoms for their study, Schroeder et al.
[14] showed that plaque
densitometry was accurate on thin-slice images (1 vs 2.5 mm), which are highly
dependent on surrounding contrast enhancement. Cademartiri et al.
[15], showing that CT
attenuation of plaque significantly varied (22 ± 22, 50 ± 26,
107 ± 36, 152 ± 67 H) according to luminal enhancement (35
± 10, 91 ± 7, 246 ± 18, 511 ± 89 H, respectively),
concluded that it is difficult to identify absolute ranges of attenuation that
relate to specific plaque characteristics.
We thought about the need for an experiment using 64-MDCT with a thin
collimation and improved temporal and spatial resolution, as pointed out by
Nikolaou et al. [16]. For such
an experiment, the CT attenuations of plaque models should mimic those of real
plaques measured on cadavers' arteries
[11,
12], and the range of coronary
enhancement should simulate that seen in clinical situations. In addition,
most important but never reported, to our knowledge, a pulsating cardiac
phantom with variable heart rates was validated with different temporal
resolutions. The purpose of this study was to validate the accuracy of 64-MDCT
densitometry of soft and intermediate plaques using coronary artery plaque
models on a pulsating cardiac phantom.
Materials and Methods
Intravascular Enhancement of Clinical Studies: Coronary CT Angiography on 64-MDCT
This part of the study was to investigate patients' Hounsfield attenuation
values for blood enhancement in clinical 64-MDCT coronary angiography and to
adjust the attenuation values for the coronary artery fluid used for the
phantom study. This study was approved by our institutional review committee.
After seven patients with heavy calcifications or a stent or stents in the
main coronary branches had been excluded, 30 consecutive patients (19 men and
11 women; mean age ± SD, 64 ± 11 years; age range, 42–85
years) undergoing cardiac MDCT investigation were enrolled. Both this clinical
study and the following phantom experiments of retrospective ECG-gated 64-MDCT
coronary angiography were performed using the same unit (LightSpeed VCT, GE
Healthcare) with the same scanning parameters except tube current and pitch.
No ß-blockers were administered in the series.
After a test injection of 15 mL of nonionic contrast medium (iopamidol
[Iopamiron 370, Schering]) was administered, a timing bolus scan was obtained
with contrast medium (amount of contrast medium = 0.7 x body weight) at
an injection time of 10 seconds, followed by a 25-mL saline chaser
administered at the same rate. The scan covered the entire volume of the heart
during a single breath-hold, consisting of 5 or 6 heartbeats, with
simultaneous recording of the ECG trace. Detector collimation was 64 x
0.625 mm, gantry rotation speed was 350 milliseconds per rotation, and tube
voltage was 120 kV at a tube current of 550–750 mAs (depending on
patient size, with a dose-reduction model using the ECG modulation
technique).
CT pitch factors ranged from 0.18 to 0.24 by heart rate according to the
manufacturer's recommendations for a coronary CT angiography protocol. For
heart rates < 75 beats per minute (bpm), a half-scan algorithm was applied;
for heart rates
75 bpm, a two-segment reconstruction algorithm, offering
improved temporal resolution, was used. For image reconstruction, cardiac
phase imaging with the center of the temporal window corresponding to 65% of
the R-R interval was used. Other cardiac phase images were reconstructed in
cases in which the image quality of the phase images was better than that at
65% of the R-R interval. CT attenuation values in the regions of interest
(ROIs) were 316 ± 67 H (range, 167–415 H), 309 ± 65 H
(range, 173–429 H), 297 ± 66 H (range, 173–420 H), and 310
± 60 H (range, 168–418 H) on the proximal left main, left
anterior descending, left circumflex, and right coronary arteries,
respectively. Thereafter, for coronary artery enhancement models, four
concentrations of contrast medium and water with CT attenuation values of 150,
250, 350, and 450 H were prepared.
Pulsating Cardiac Phantom
A prototype cardiac phantom was used (ALPHA 2, Fuyo Corporation). The
phantom consists of five components: driver, control, support, rubber balloon,
and ECG (Fig. 1). A controller
with an ECG-synchronizer drives the balloon. The construction of the phantom
is described in detail elsewhere
[17,
18]. The main characteristic
features of this phantom are programmable variable heart rate sequences and
programmable heart movements that mimic natural heart movements. In this
study, eight heart rate sequences were programmed
(Fig. 2).
Coronary Artery Plaque Models
Acrylonitrile–butadiene–styrene resin (47 H) and acrylic (110
H) were used to represent soft and intermediate plaques, respectively. In
addition, three plaque shapes (D-shaped, centric, and eccentric) and two
coronary diameters (3 and 4 mm) were prepared. Using different combinations of
plaque shape and artery diameter, the manufacturer of the phantom made seven
plaque models. All plaque models had three levels of stenosis
(Fig. 3). These seven plaque
models were attached to the balloon phantom (mimicking the heart), with the
long axis of the model corresponding to the z-axis, and were
surrounded by water (Fig.
4).

View larger version (6K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 3 —Coronary artery plaque models. This drawing shows plaque
size, plaque shape, and stenosis levels of coronary artery plaque models.
Seven plaques were prepared with different combinations of plaque shape,
plaque CT attenuation, and coronary artery diameter, respectively: D-shaped,
40 H, 4 mm; D-shaped, 40 H, 3 mm; centric, 40 H, 4 mm; centric, 40 H, 3 mm;
eccentric, 40 H, 4 mm; D-shaped, 100 H, 4 mm; and centric, 100 H, 4 mm.
|
|

View larger version (97K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 4 —Balloon phantom. Seven plaque models were attached to balloon
phantom (mimicking heart) and were surrounded by water. Balloon was filled
with mixture of water and contrast medium (CT attenuation = 40 H).
|
|
Cardiac Phantom Scan
Volumetric data of the phantom were obtained. The tube current used was 650
mA. The pitch was set to 0.175 to allow improved temporal resolution by two-
and four-segment reconstruction algorithms. Other scanning parameters were the
same as those used for the clinical study, as described earlier. A half-scan
algorithm provided a temporal resolution of 175 milliseconds; a two-segment,
97–137 milliseconds; and a four-segment, 52–95 milliseconds. The
temporal resolution of a two- or four-segment reconstruction algorithm varied
on sequences performed with heart rate shifting.
Theme 1: Heart Rate and Temporal Resolution for the Imaging of Coronary Artery Plaque
Degradation of image quality caused by motion artifacts decreased the
detection of coronary artery plaque and the accuracy of CT densitometry. Image
quality at a static state and image quality at eight heart rates were
compared. A two-segment reconstruction algorithm was applied for heart rates
of
65 bpm and a four-segment algorithm was used for heart rates of
80 bpm. The segmentation algorithm was not used at static state or at 50 bpm
because of its lack of effectiveness in improving temporal resolution. For
intracoronary artery enhancement, 250 and 350 H, which are considered
representative levels judging from our clinical study results, were chosen.
Details of the settings are summarized in Appendix 1. The plaque used for this
theme was a D-shaped model (semicircular shape) with 50% stenosis.
Image quality was subjectively evaluated by three reviewers (experience
interpreting cardiac CT: 7, 3, and 3 years) who were unaware of the kinds of
plaque models and who graded image quality using a 3-point scale: 2, no or
minor motion artifacts, recognized as semicircular shape; 1, mild motion
artifacts, however recognized as almost semicircular shape; 0, significant
motion artifacts, not recognized as semicircular shape. When the grades
assigned for image quality differed among reviewers, the consensus of two
reviewers was defined as the grade. Images were displayed with fixed window
settings (window width, 700 H; window level, 200 H).
Theme 2: Does CT Attenuation of Coronary Plaque Change by Coronary Artery Enhancement?
CT attenuation values of soft and intermediate plaques were compared at
four different intracoronary artery enhancement values (150, 250, 350, and 450
H). ROIs that were 1 mm2 were set at the center of the plaques
(Fig. 5A). Static and
low-heart-rate sequences were chosen to avoid or minimize the effect of motion
artifacts. Details of the settings are summarized in Appendix 2. CT
densitometry was performed in five slices per plaque in each circumstance by
one reviewer (3 years of experience).

View larger version (14K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 5A —Region-of-interest (ROI) setting in coronary artery plaque.
Drawing shows typical ROI placement (dotted lines) in coronary artery
plaques (gray). All ROIs were ovoid and 1 mm2. ROIs were
set at center of plaques in theme 2 (A), near arterial lumen and at
center of plaque in theme 3 (B), at center of plaques in three stenotic
levels in theme 4 (C), and at center of plaques in various types of
plaque shape and coronary artery diameters in theme 5 (D).
|
|
Theme 3: Does Placing the ROI near the Arterial Lumen Affect CT Attenuation?
ROIs were placed near the arterial lumen and at the center of the plaque
(Fig. 5B), and the CT
attenuation values of soft and intermediate plaques were compared between
measurement sites. Two different CT attenuation values of intracoronary artery
enhancement (250 and 350 H), which were representative in our clinical study,
were used. Details of the settings are summarized in Appendix 3.

View larger version (10K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 5B —Region-of-interest (ROI) setting in coronary artery plaque.
Drawing shows typical ROI placement (dotted lines) in coronary artery
plaques (gray). All ROIs were ovoid and 1 mm2. ROIs were
set at center of plaques in theme 2 (A), near arterial lumen and at
center of plaque in theme 3 (B), at center of plaques in three stenotic
levels in theme 4 (C), and at center of plaques in various types of
plaque shape and coronary artery diameters in theme 5 (D).
|
|
Theme 4: Is CT Attenuation of Coronary Plaque Affected by the Level of Arterial Stenosis?
CT attenuation values of soft and intermediate plaques were compared at
three levels of stenosis: 25%, 50%, and 75% of the arterial diameter
(Fig. 5C). Details of the
settings are summarized in Appendix 4.

View larger version (9K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 5C —Region-of-interest (ROI) setting in coronary artery plaque.
Drawing shows typical ROI placement (dotted lines) in coronary artery
plaques (gray). All ROIs were ovoid and 1 mm2. ROIs were
set at center of plaques in theme 2 (A), near arterial lumen and at
center of plaque in theme 3 (B), at center of plaques in three stenotic
levels in theme 4 (C), and at center of plaques in various types of
plaque shape and coronary artery diameters in theme 5 (D).
|
|
Theme 5: Is CT Attenuation of Coronary Plaque Affected by the Coronary Artery Diameter or the Stenosis Shape?
CT attenuation values of soft plaque were compared for the following five
combinations of plaque shape and coronary artery diameter: D-shaped and 3 mm,
D-shaped and 4 mm, centric and 3 mm, centric and 4 mm, and eccentric and 4 mm.
The stenotic ratio was 50% in area. ROIs were set at the center of the plaques
(Fig. 5D). Details of the
settings are summarized in Appendix 5.

View larger version (9K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 5D —Region-of-interest (ROI) setting in coronary artery plaque.
Drawing shows typical ROI placement (dotted lines) in coronary artery
plaques (gray). All ROIs were ovoid and 1 mm2. ROIs were
set at center of plaques in theme 2 (A), near arterial lumen and at
center of plaque in theme 3 (B), at center of plaques in three stenotic
levels in theme 4 (C), and at center of plaques in various types of
plaque shape and coronary artery diameters in theme 5 (D).
|
|
Statistical Analysis
All statistical analyses were performed using a commercially available
software package (Statcel2, OMS Publishing). For statistical analyses,
Mann-Whitney U, repeated measures analysis-of-variance, Wilcoxon's
signed rank, and Kruskal-Wallis tests were used to determine differences. When
statistical significance was observed by repeated measures analysis of
variance, the results were made post hoc using the Scheffé test for
multiple pairwise comparisons. A p value of < 0.05 was considered
to identify significant differences.
Results
Theme 1
Subjective evaluation of image quality of plaques for different
combinations of heart rate sequence and reconstruction algorithm is summarized
in Table 1. Assessable image
quality of soft plaque was obtained with all combinations of heart rates and
reconstruction algorithms tested, whereas images of intermediate plaque were
of poor quality on a coronary artery enhancement level of 250 H at
high-heart-rate sequences. The image quality of intermediate plaques at 350 H
was better than that at 250 H (Mann-Whitney U test, p <
0.01). When reconstruction algorithms were compared, the segmentation
algorithm with improved temporal resolution tended to yield better image
quality.
View this table:
[in this window]
[in a new window]
|
TABLE 1: Subjective Evaluation of Image Quality with Different Combinations of
Heart Rate Sequences and Reconstruction Algorithms
|
|
Theme 2
CT attenuation values of soft and intermediate plaques on four levels of
intracoronary artery enhancement for four combinations of heart rate and
reconstruction algorithm are shown in Figure
6A,
6B. CT attenuation values of
soft plaque were overestimated on intracoronary artery enhancement levels of
350 and 450 H (p < 0.01). The CT densitometry measurements were
accurate and lower for the static model and at 50 bpm (p < 0.01).
Intermediate plaques were not detectable on intracoronary artery enhancement
of 150 H.

View larger version (13K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 6A —Relationship of CT attenuation values between plaque and
intracoronary enhancement. Graphs show results for soft (A) and
intermediate (B) plaques. Repeated measures analysis of variance
revealed that CT attenuation values of soft plaque were different among
intracoronary artery enhancement levels (p < 0.01) and
combinations of heart rates and reconstruction algorithms (p <
0.01). Scheffé test for multiple pairwise comparisons revealed that CT
attenuation values were significantly different between intracoronary
enhancement of 150 and 350 H (p < 0.01), 150 and 450 H (p
< 0.01), and 250 and 450 H (p < 0.01) on static cardiac phantom
using half reconstruction and between intracoronary enhancement of 150 and 350
H (p < 0.01), 150 and 450 H (p < 0.01), 250 and 350 H
(p < 0.01), and 250 and 450 H (p < 0.01) on cardiac
phantom at 50 beats per minute (bpm) using half reconstruction algorithm.
Scheffé test revealed that CT attenuation values of plaque on
intracoronary artery enhancement of 150 H were significantly different between
static cardiac phantom with half reconstruction algorithm and 65 bpm with half
reconstruction (p < 0.01) and between 50 bpm with half
reconstruction and 65 bpm with half reconstruction (p < 0.01).
Scheffé test also revealed that CT attenuation values of plaque on
intracoronary artery enhancement of 250 H were significantly different between
static phantom with half reconstruction and 65 bpm with half reconstruction
(p < 0.01) and between 50 bpm with half reconstruction and 65 bpm
with half reconstruction (p < 0.01). In contrast, CT attenuation
values of intermediate plaque were not statistically different based on
intracoronary artery enhancement level (p = 0.09) or combinations of
heart rate and reconstruction algorithm (p = 0.10).
|
|

View larger version (13K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 6B —Relationship of CT attenuation values between plaque and
intracoronary enhancement. Graphs show results for soft (A) and
intermediate (B) plaques. Repeated measures analysis of variance
revealed that CT attenuation values of soft plaque were different among
intracoronary artery enhancement levels (p < 0.01) and
combinations of heart rates and reconstruction algorithms (p <
0.01). Scheffé test for multiple pairwise comparisons revealed that CT
attenuation values were significantly different between intracoronary
enhancement of 150 and 350 H (p < 0.01), 150 and 450 H (p
< 0.01), and 250 and 450 H (p < 0.01) on static cardiac phantom
using half reconstruction and between intracoronary enhancement of 150 and 350
H (p < 0.01), 150 and 450 H (p < 0.01), 250 and 350 H
(p < 0.01), and 250 and 450 H (p < 0.01) on cardiac
phantom at 50 beats per minute (bpm) using half reconstruction algorithm.
Scheffé test revealed that CT attenuation values of plaque on
intracoronary artery enhancement of 150 H were significantly different between
static cardiac phantom with half reconstruction algorithm and 65 bpm with half
reconstruction (p < 0.01) and between 50 bpm with half
reconstruction and 65 bpm with half reconstruction (p < 0.01).
Scheffé test also revealed that CT attenuation values of plaque on
intracoronary artery enhancement of 250 H were significantly different between
static phantom with half reconstruction and 65 bpm with half reconstruction
(p < 0.01) and between 50 bpm with half reconstruction and 65 bpm
with half reconstruction (p < 0.01). In contrast, CT attenuation
values of intermediate plaque were not statistically different based on
intracoronary artery enhancement level (p = 0.09) or combinations of
heart rate and reconstruction algorithm (p = 0.10).
|
|
Theme 3
Plaque CT attenuation values near the arterial lumen and at the center of
plaque are shown in Figure 7.
In all combinations of plaque (soft and intermediate) and intracoronary
enhancement (250 and 350 H), plaque CT attenuation values near the lumen were
overestimated and higher than those at the center (p < 0.01).

View larger version (15K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 7 —Differences in CT attenuation values between
region-of-interest placement inside plaque. CT attenuation values of 250 and
350 H were chosen as representative levels from our clinical study results.
Wilcoxon's signed rank test revealed that CT attenuation values of plaque near
lumen (dark gray) were overestimated and higher than those at center
(light gray): soft plaque and intracoronary enhancement of 250 H,
soft plaque and 350 H, intermediate plaque and 250 H, and intermediate plaque
and 350 H (p < 0.01).
|
|
Theme 4
Plaque CT attenuation values in three stenotic levels are shown in
Figure 8. For all combinations
of plaque (soft and intermediate) and intracoronary artery enhancement (250
and 350 H), CT attenuation values with 25% stenosis were overestimated and
higher than those with 50% or 75% stenosis (p < 0.01).

View larger version (14K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 8 —Relationship between plaque CT attenuation values and
stenosis level. CT attenuation values of plaque between three stenotic
levels—that is, 25% (dark gray), 50% (light gray), and
75% (black) (in diameter). Kruskal-Wallis test revealed that CT
attenuation values of plaque were different between stenosis levels: soft
plaque and intracoronary enhancement of 250 H, soft plaque and 350 H,
intermediate plaque and 250 H, and intermediate plaque and 350 H (p
< 0.01).
|
|
Theme 5
Plaque CT attenuation values in different combinations of shape and
coronary artery diameter are shown in
Figure 9. Plaque CT
attenuation values were different between the combinations (p <
0.01). D-shaped plaque in a coronary artery with a diameter of 4 mm showed
lower CT attenuation values than the other combinations and was close to the
real CT attenuation.

View larger version (17K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 9 —Plaque CT attenuation values in combinations of plaque shape
and coronary artery diameter. Kruskal-Wallis test revealed that CT attenuation
values of plaque were statistically different between combinations of soft
plaque and coronary artery enhancement of 250 H (gray) and soft
plaque and coronary artery enhancement of 350 H (black) (p
< 0.01).
|
|
Discussion
Although the identification of calcified plaque is straightforward because
of its higher CT attenuation, differentiation of noncalcified plaque from
calcified plaque is challenging. A lipid core larger than 1 mm2 or
a lipid-core-to-plaque ratio of greater than 20% and a fibrous cap thinner
than 0.7 mm have been shown to correlate well with plaque rupture
[19]. The usual location of a
lipid core within an atherosclerotic plaque is just below the thin cap.
Factors and Circumstances Necessary for Accurate CT Densitometry of Coronary Plaque
There is a consensus that a low heart rate is an advantage for cardiac CT
examination and that multisegment reconstruction, with its improved temporal
resolution, is effective in high heart rates, especially when stable. In the
current study, the two- or four-segment reconstruction was sometimes helpful
for better delineation of the plaques. However, because of the finding that
soft plaque showed higher CT attenuation values on half- and two-segment
reconstructions at 65 bpm, plaque CT densitometry seems more accurate and can
be performed better on a patient with a lower heart rate.
The results of the current study suggest that coronary artery enhancement
has a significant impact on the CT attenuation of plaque, especially that near
the coronary artery lumen—that is, just below the thin cap. This is
considered due to partial volume averaging, which is theoretically dominant
for low-attenuation plaque surrounded by the high-attenuation coronary artery
lumen. The results suggest that it is difficult to predict the presence of
soft plaque on the basis of a CT attenuation measured in an ROI set near the
coronary lumen. High coronary artery enhancement (350 H) is advantageous for
better delineation of both soft and intermediate plaques. However, as shown in
this study, soft plaque (50% stenosis) was detected on all heart rate
sequences on coronary artery enhancement of 250 H. This finding suggests that
this level of enhancement might be better for characterization of soft plaque,
as suggested by Schroeder et al.
[13,
14].
From the results of themes 4 and 5, it seems that a minimal plaque
size—for example, resulting in luminal stenosis of 50% or more— is
needed to precisely measure the attenuation of plaque, especially of soft
plaque. An appropriate size for an ROI remains unresolved. Because CT
densitometry of D-shaped soft plaque causing 50% stenosis within a 3-mm
coronary artery failed in yielding accurate measures, it follows that an ROI
should be situated at the center of plaque to avoid the influence of coronary
artery enhancement.
Technical matters for enhancement of the left ventricle remain to be
mentioned. We filled a mixture of water and contrast medium (40 H) into a
balloon (mimicking the heart) instead of filling a mixture with the same
enhancement as the coronary arteries. In actual coronary CT angiography, the
left ventricle is enhanced to the same level as the coronary arteries, whereas
the enhancement of the right ventricle and atrium has already decreased. The
wall of the left ventricle does not enhance much at the timing of coronary CT
angiography. In addition to these technical matters, the epicardial coronary
arteries are some distance from the heart. We therefore think that the CT
attenuation of coronary plaque is not influenced much by the enhancement of
the heart except in two instances: The plaque in the coronary orifice is
influenced by the enhancement of the aorta, and the plaque in the myocardial
bridge is influenced by the enhancement of the left ventricle.
Study Limitations
This study has several limitations. Atherosclerotic plaques have a complex
composition—that is, lipid-rich, fibrous, and calcified areas coexist
and are often intermixed. We prepared only one CT attenuation as a model for
each of the soft (47 H) and the intermediate (110 H) plaques, although the
previously reported CT attenuation values of the plaques had various ranges.
We used plaques with a uniform CT attenuation because the purpose of this
study was to investigate the effect of coronary artery enhancement on the CT
attenuation of plaque. In addition, the shape of the coronary artery was
always round; therefore, positive remodeling was not simulated. Next, we
changed coronary artery enhancement levels on static and low-heart-rate
sequences and did not investigate coronary artery enhancement on
high-heart-rate sequences. This omission needs to be mentioned if the results
are translated into an in vivo situation. Finally, we did not simulate large
variations of heart rate, such as arrhythmia or premature heart beat, or
changes in body posture.
In conclusion, coronary artery enhancement has a significant impact on
64-MDCT densitometry of coronary artery plaque, especially of soft plaque.
Large plaque size, densitometry performed not near the lumen but at the center
of the plaque, intracoronary artery enhancement of 250 H, and low heart rates
increase the accuracy of plaque densitometry.




View this table:
[in this window]
[in a new window]
|
APPENDIX 3: Design for Theme 3: Does Placing the Region of Interest (ROI) near the
Arterial Lumen Affect CT Attenuation?
|
|
View this table:
[in this window]
[in a new window]
|
APPENDIX 5: Design for Theme 5: Is CT Attenuation of Coronary Plaque Affected by the
Coronary Artery Diameter or the Stenosis Shape?
|
|
References
- Virmani R, Burke AP, Farb A, Kolodgie FD. Pathology of the
vulnerable plaque. J Am Coll Cardiol2006; 47[suppl 8]:C13
–C18[Abstract/Free Full Text]
- Naghavi M, Libby P, Falk E, et al. From vulnerable plaque to
vulnerable patient: a call for new definitions and risk assessment strategies:
part I. Circulation 2003;108
:1664
–1672[Abstract/Free Full Text]
- Nissen SE, Yock P. Intravascular ultrasound: novel
pathophysiological insights and current clinical applications.
Circulation 2001;103
: 604–616[Abstract/Free Full Text]
- Patwari P, Weissman NJ, Boppart SA, et al. Assessment of coronary
plaque with optical coherence tomography and high-frequency ultrasound.
Am J Cardiol 2000;85
: 641–644[CrossRef][Medline]
- Stefanadis C, Diamantopoulos L, Vlachopoulos C, et al. Thermal
heterogeneity within human atherosclerotic coronary arteries detected in vivo:
a new method of detection by application of a special thermography catheter.
Circulation 1999;99
:1965
–1971[Abstract/Free Full Text]
- Takano M, Mizuno K, Okamatsu K, Yokoyama S, Ohba T, Sakai S.
Mechanical and structural characteristics of vulnerable plaques: analysis by
coronary angioscopy and intravascular ultrasound. J Am Coll
Cardiol 2001; 38:99
–104[Abstract/Free Full Text]
- Kopp AF, Schroeder S, Baumbach A, et al. Non-invasive
characterisation of coronary lesion morphology and composition by multislice
CT: first results in comparison with intracoronary ultrasound. Eur
Radiol 2001; 11:1607
–1611[CrossRef][Medline]
- Schroeder S, Kopp AF, Baumbach A, et al. Noninvasive detection and
evaluation of atherosclerotic coronary plaques with multislice computed
tomography. J Am Coll Cardiol 2001;37
:1430
–1435[Abstract/Free Full Text]
- Leber AW, Knez A, Becker A, et al. Accuracy of multidetector spiral
computed tomography in identifying and differentiating the composition of
coronary atherosclerotic plaques: a comparative study with intracoronary
ultrasound. J Am Coll Cardiol 2004;43
:1241
–1247[Abstract/Free Full Text]
- Pohle K, Achenbach S, MacNeill B, et al. Characterization of
non-calcified coronary atherosclerotic plaque by multi-detector row CT:
comparison to IVUS. Atherosclerosis 2007;190
: 174–180[CrossRef][Medline]
- Estes JM, Quist WC, Lo Gerfo FW, et al. Noninvasive
characterization of plaque morphology using helical computed tomography.
J Cardiovasc Surg (Torino) 1998;39
: 527–534[Medline]
- Becker CR, Nikolaou K, Muders M, et al. Ex vivo coronary
atherosclerotic plaque characterization with multi-detector-row CT.
Eur Radiol 2003;13
:2094
–2098[CrossRef][Medline]
- Schroeder S, Kuettner A, Leitritz M, et al. Reliability of
differentiating human coronary plaque morphology using contrast-enhanced
multislice spiral computed tomography: a comparison with histology.
J Comput Assist Tomogr 2004;28
: 449–454[CrossRef][Medline]
- Schroeder S, Flohr T, Kopp AF, et al. Accuracy of density
measurements within plaques located in artificial coronary arteries by X-ray
multislice CT: results of a phantom study. J Comput Assist
Tomogr 2001; 25:900
–906[CrossRef][Medline]
- Cademartiri F, Mollet NR, Runza G, et al. Influence of
intracoronary attenuation on coronary plaque measurements using multislice
computed tomography: observations in an ex vivo model of coronary computed
tomography angiography. Eur Radiol 2005;15
:1426
–1431[CrossRef][Medline]
- Nikolaou K, Flohr T, Knez A, et al. Advances in cardiac CT imaging:
64-slice scanner. Int J Cardiovasc Imaging2004; 20:535
–540[CrossRef][Medline]
- Horiguchi J, Shen Y, Akiyama Y, et al. Electron beam CT versus
16-MDCT on the variability of repeated coronary artery calcium measurements in
a variable heart rate phantom. AJR 2005;185
:995
–1000[Abstract/Free Full Text]
- Horiguchi J, Shen Y, Akiyama Y, et al. Electron beam CT versus
16-slice spiral CT: how accurately can we measure coronary artery calcium
volume? Eur Radiol 2006;16
: 374–380[CrossRef][Medline]
- Ge J, Chirillo F, Schwedtmann J, et al. Screening of ruptured
plaques in patients with coronary artery disease by intravascular ultrasound.
Heart 1999; 81:621
–627[Abstract/Free Full Text]

CiteULike
Complore
Connotea
Del.icio.us
Digg
Reddit
Technorati What's this?